Standard parameters: | 3/5-Lead ECG, ECG, NIBP, BMV SpO2, HR/PR, 2-RESP, 2-TEMP, Rechargeable Battery |
Optional configuration: | Built-in printer, BMV SpO2, Nellcor SpO2, Masimo SpO2, 2-IBP, ETCO2(mainstream or side), WLAN accessory, Wall Mount, Moving cart |
ECG | |
Number of Leads: | 3 or 5 leads Lead View |
Lead output: | 3 wires/5 wires ECG cable |
User Selectable: | I, II, III, aVR, aVL, aVF, V (5 Lead); I, II or III (3 lead) |
Gain Selection: | x1/4, x1/2, x1 and x2 Frequency Response |
Diagnostic: | 0.05 to 90HZ |
Monitor: | 0.5 to 35 HZ |
Surgery: | Surgery: 1-15HZ |
Calibration Signal: | 1 (mV p-p) |
Accuracy : | ECG calibration: ± 1 mv Heart rate accuracy: ± 1% |
ECG Signal Range: | ±8 m V (Vp-p) |
Review | Available |
NIBP (Non-invasive Blood Pressure) | |
Technique: | Oscillometric during inflation range |
Measurement type: | Adult, pediatric, neonate |
Measurement parameter: | Systolic, Diastolic, Mean |
Adult: | 40-270 mmHg |
Pediatric: | 40-240 mmHg |
Neonate: | 10-150 mmHg |
Measurement cycle: | <40 sec, typical |
Cycles (Selectable): | 2, 3, 5,10,15, 30 min / 1, 2, 4hr |
STAT model: | 5 minutes of continuous reading Max Allowable Cuff Pressure |
Adult: | 300 mmHg |
Pediatric: | 240 mmHg |
Neonate: | 150 mmHg |
Resolutions: | 1 mmHg |
Transducer Accuracy : | ±2 mmHg or ±2% of reading |
SpO2 | |
Display type: | Waveform, data |
Measurement range: | 0~100% |
Resolution: | 1 BPM |
Pulse rate range: | 30~250bpm |
Accuracy: | ±1bpm |
CO2 (option) | |
Measurement range: | 0% ~ 10.0% (Carbon dioxide inhalation and carbon dioxide exhalation) |
Accuracy: | ±2mmHg (<5% Measured values) 10%(>5% Measured values) |
Pulse Oximetry | |
Range: | 0 to 100% |
Resolution: | 1% |
Accuracy: | 70% to 99% range ±2%,1% to 69% range ±5% |
Bodymark: | •Over 144 bodymark patterns for all applications •Support transducer mark |
Method: | Dual wavelength LED |
Respiration Rate: | |
Rate: | Adult: 0 ~ 120 rpm / Neo/Ped 0 ~ 150 rpm |
Resolution: | 1 rpm |
Accuracy: | ± 2 breath/min |
Heart (Pulse) Rate | |
Range: | Adult 15-300 bpm, Pediatric 15-350 bpm |
Accuracy: | ±2bpm or ±3% whichever is greater |
Trends | |
Memory: | 96 hr at 30 second intervals for a NIBP reading every 5 minutes |
Display : | Trabular; Graphical |